4 brc - control parameters, Control parameters – Metrohm tiamo 2.1 Manual User Manual
Page 805
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.1
■■■■■■■■
791
Max. waiting time
If signal drift has been switched off or has not yet been reached, then the
measured value will be accepted when the maximum waiting time has
elapsed.
Range
0 ... 999999 s
Default value
15 s
Pause
Pause
Waiting time before the start of the titration.
Range
0 ... 999999 s
Default value
0 s
5.6.3.7.4
BRC - Control parameters
Tab: Method
▶ BRC ▶ Properties... ▶ Control parameters
Command name
Name of the command.
Input
25 characters
Endpoint
Ep with
Measured value for the endpoint.
Range
-1250.0 ... 1250.0 mV
Default value
200.0 mV
Titration rate
Titration rate
The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for Control are not displayed.
The User setting must be selected in order to be able to edit these param-
eters.
Selection
slow | optimal | fast | User
Default value
optimal
slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwanted
EPs.