KFC - Stop conditions
Tab: Method
▶ KFC ▶ Properties... ▶ Stop conditions
Command name
Name of the command.
Input
25 characters
The conditions for stopping the titration are defined on this tab.
Stop time
Stops when the preset time has elapsed since the start of the titration. off
means no stop.
Range
1 ... 999999 s
Selection
off
Default value
off
5.6.3.6.7
KFC - Conditioning
Tab: Method
▶ KFC ▶ Properties... ▶ Conditioning
Command name
Name of the command.
Input
25 characters
Conditioning
on | off (Default value: on)
Enable/Disable the conditioning. If the conditioning is disabled, the following
parameters are not visible.
Start drift
The determination can only be started when the currently measured drift is
less than the start drift entered here. In such cases, the Status in the