6.7 Subwindow Sensors
■■■■■■■■■■■■■■■■■■■■■■
1278
■■■■■■■■
tiamo 2.1
Range
-2.147E+9 ... 2.147E+9
Default value
7.500
Intervention limits for pH(0) (only for pH electrodes)
on | off (Default value: off)
If this option is switched on, then the values for pH(0) will be shown in
red if the limits are exceeded.
Lower limit
Lower intervention limit for pH(0).
Range
-2.147E+9 ... 2.147E+9
Default value
6.000
Upper limit
Upper intervention limit for pH(0).
Range
-2.147E+9 ... 2.147E+9
Default value
8.000
Warning limits for E(0) (only for ISE electrodes)
on | off (Default value: off)
If this option is switched on, then the values for E(0) will be shown in
orange if the limits are exceeded.
Lower limit
Lower warning limit for E(0).
Range
-2.147E+9 ... 2.147E+9 mV
Default value
-15 mV
Upper limit
Upper warning limit for E(0).
Range
-2.147E+9 ... 2.147E+9 mV
Default value
15 mV
Intervention limits for E(0) (only for ISE electrodes)
on | off (Default value: off)
If this option is switched on, then the values for E(0) will be shown in
red if the limits are exceeded.
Lower limit
Lower intervention limit for E(0).