Metrohm tiamo 2.1 Manual User Manual
Page 351
■■■■■■■■■■■■■■■■■■■■■■
4 Database
tiamo 2.1
■■■■■■■■
337
Measuring input:
Number of the measuring input to which the sensor is connected (+ desig-
nation iConnect for intelligent sensors).
Slope:
Electrode slope of the sensor used (in % for pH sensor or in mV for ISE).
E (0):
Electrode zero point of the sensor used (in mV only for ISE).
pH (0):
Electrode zero point of the sensor used (dimensionless for pH sensor).
c (blank):
Blank value of the ISE sensor used.
Cell constant:
Cell constant of the conductivity sensor used.
Calibration temperature:
Temperature during the calibration.
Calibration date:
Temperature during the calibration.
Calibration method:
Shows the method with which the calibration was carried out.
User:
Shows the user logged in during the calibration or who entered the calibra-
tion data manually.
Measuring input
Shows the measuring input with which the calibration was carried out.
Common variable 'Name'
Information about the common variable used.
Value:
Value and unit of the common variables after the last value assignment.
Assignment date:
Date and time of last value assignment.