7 met upol - additional evaluations, Additional evaluations – Metrohm tiamo 2.1 Manual User Manual
Page 645

■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.1
■■■■■■■■
631
descending
Only equivalence points with a negative slope of the titration curve will
be recognized.
off
Equivalence point recognition is switched off.
Evaluation with measured value window (Upol)
EP criterion
Criterion for the recognition of equivalence points which is valid for all win-
dows. Equivalence points whose ERC is smaller than the set EP criterion will
not be recognized.
Titrino
Range
0.1 ... 99.9 µA
Default value
2.0 µA
[New]
Opens the dialog window Measured value # in which the parameters for
a new window can be entered (see Chapter 5.6.3.3.5.9, page 637).
[Properties]
Opens the dialog window Measured value # in which the parameters for
the selected window can be edited (see Chapter 5.6.3.3.5.9, page 637).
[Delete]
Deletes the window selected in the table.
Evaluation with volume window (mL)
[New]
Opens the dialog window Volume window # in which the parameters for
a new window can be entered (see Chapter 5.6.3.3.5.10, page 638).
[Properties]
Opens the dialog window Volume window # in which the parameters for
the selected window can be edited (see Chapter 5.6.3.3.5.10, page
638).
[Delete]
Deletes the window selected in the table.
5.6.3.3.5.7
MET Upol - Additional evaluations
Tab: Method
▶ MET Upol ▶ Properties... ▶ Additional evaluations
Command name
Name of the command.
Input
25 characters