5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

530

■■■■■■■■

tiamo 2.1

Window = Time
Range

0 ... 999999 s

Default value

0 s

Upper limit

Value for the upper limit of the window.

Window = Measured value
Range

-200.0 ... 200.0 µA

Default value

200.0 µA

Window = Volume
Range

0.00000 ... 9999.99 mL

Default value

9999.99 mL

Window = Time
Range

0 ... 999999 s

Default value

999999 s

Gran evaluation

Note

This evaluation method is possible only with Titrando and 855.

Gran evaluation

on | off (Default value: off)

If this option is enabled, then titration curves will be evaluated in accordance
with the Gran Plot procedure(see Chapter 5.6.3.9.4, page 856).

Initial volume

Volume which is present in the measuring vessel before the command is
started.

Range

0.01 ... 9999.99 mL

Default value

50.00 mL

Lower limit

Value for the lower limit of the Gran evaluation range.

Range

-200.0 ... 200.0 µA

Default value

-200.0 µA