5 Method
tiamo 2.1
■■■■■■■■
777
Ipol
10 µA
20 µA
30 µA
EP
50 mV
100 mV
150 mV
Dynamics
70 mV
100 mV
120 mV
Max. rate
Maximum rate with which iodine is generated outside the Control range.
Range
1.5 ... 2240.0 µg/min
Default value
1000.0 (slow), maximum (optimal), maximum
(fast) µg/min
Selection
maximum
Min. rate
This parameter defines the minimum rate with which iodine is generated
inside the Control range. The smaller the rate the slower the titration and
the higher the accuracy.
756, 831
Range
0.3 ... 999.9 µg/min
Default value
0.3 (slow), 15.0 (optimal), 30.0 (fast) µg/min
Selection
minimum
851, 852
Range
0.3 ... 999.0 µg/min
Default value
0.3 (slow), 15.0 (optimal), 30.0 (fast) µg/min
Selection
minimum
Stop criterion
Stop criterion
Switches off titration when the endpoint has been reached and the stop
criterion has been met.
756, 831
Selection
drift | rel. drift
Default value
rel. drift
851, 852
Selection
drift | rel. drift | time | off
Default value
rel. drift
Stop drift
The titration is stopped when the endpoint and the stop drift have been
reached. This parameter is only shown for Stop criterion = drift.