5 Method

tiamo 2.1

■■■■■■■■

917

Stop meas. value

Stops when the preset measured value has been reached since the start of
the measurement.

Titrando, 855, 867
Range

-20.0 ... 150.0 °C

Selection

off

Default value

off

off
No stop.

Measurement without drift control
If the option Measurement without drift control has been selected, then
the following parameters will be displayed:

Measuring time

Maximum period of time for measurement.

Titrando, 855, 867
Range

0 ... 999999 s

Default value

120 s

Titrino
Range

0 ... 9999 s

Default value

120 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.

Titrando, 855, 867
Range

0.1 ... 999999.0 s (Increment: 0.1)

Default value

2.0 s

Titrino
Range

0.08 ... 16200 s (Increment: 0.08)

Default value

2.0 s

Stop meas. value

Stops when the preset measured value has been reached since the start of
the measurement.

Titrando, 855, 867
Range

-20.0 ... 150.0 °C

Selection

off

Default value

off