5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
866
■■■■■■■■
tiamo 2.1
Max. waiting time
If the signal drift has not yet been reached, then the measured value will be
accepted when the maximum waiting time has elapsed. If the waiting time
has not been newly entered then a waiting time that is suitable for the drift
will be calculated automatically according to the following equation:
Titrando, 855, 867
Range
0 ... 999999 s
Default value
52 s
Titrino
Range
0 ... 9999 s
Default value
52 s
Measuring interval
Time interval for entering a measuring point in the measuring point list.
Titrando, 855, 867
Range
0.1 ... 999999.0 s (Increment: 0.1)
Default value
2.0 s
Titrino
Range
0.08 ... 16200 s (Increment: 0.08)
Default value
2.0 s
Stop measured value pH
Stops when the preset measured value has been reached since the start of
the measurement.
Titrando, 855, 867
Range
-20.000 ... 20.000
Selection
off
Default value
off
off
No stop.
Measurement without drift control
If the option Measurement without drift control has been selected, then
the following parameters will be displayed:
Measuring time
Maximum period of time for measurement.