Care and use manual – Waters Envirogel GPC Cleanup Column User Manual
Page 6

[ Care and Use ManUal ]
6
Waters Envirogel GPC Cleanup Column
v. Care anD maintenanCe
General care and maintenance procedures are addressed in the Waters GPC Cleanup System Operator’s Guide. Additional symptoms that you
may observe are listed in Table 1. Recalibrate columns after any cleaning procedure.
Table 1. Troubleshooting
Problem
Possible Cause
Corrective Action
High pressure shutdown
Clogged fluid path
Change all filters in fluid path. Filter all samples and
solvents before use. Replace all tubing.
Flush column at 2 m by reversing direction of flow through
column.
Replace column.
Low operating pressure
Leaks
Tighten all fluid connections.
Insufficient solvent
Refill solvent reservoir.
Air in-line
Purge system.
Retention times too long
Pump flow rate set too low
Check flow setting and verify output.
Leaks
Tighten fluid connections.
Wrong columns
Verify length and diameter.
Retention times too short
Pump flow rate set too fast
Check flow rate setting and verify output.
Data collection device improperly connected Ensure data collection starts at inject signal.
Wrong columns
Verify length and diameter.
No peaks, poor recovery
Solvent flow too low
Verify pump pressure.
Injector blocked or leaking
Check for leaks or blockage.
Detector malfunctioning
Check wavelength, sensitivity, signal output cables.
Broad, smeared peaks, or peaks
appear as doublet
Rearrangement of packed bed or void in the
column packing
Check system performance with each column separately.
Reverse flow through affected column using 2 mL injections
of toluene or xylene at 2 mL/min. If not corrected after 10
injections, replace column.
Blockage or leak in injector sample loop
Check for leaks or excessive pressure drop across sample
loop. Back flush loop.
In-line or inlet filter is partially clogged
Reverse flow through column and inject 2 mL of
tetrahydrofuran, butylchloride, or a mixture of 1:1
cyclohexane:methylene chloride. Change the inlet filters.
Retention times shift more than
5%
Change in flow rate
Check flow rate.
Change in laboratory temperature
Stabilize laboratory temperature or column temperature.