v. Care anD maintenanCe

General care and maintenance procedures are addressed in the Waters GPC Cleanup System Operator’s Guide. Additional symptoms that you
may observe are listed in Table 1. Recalibrate columns after any cleaning procedure.

Table 1. Troubleshooting

Problem

Possible Cause

Corrective Action

High pressure shutdown

Clogged fluid path

Change all filters in fluid path. Filter all samples and
solvents before use. Replace all tubing.

Flush column at 2 m by reversing direction of flow through
column.
Replace column.

Low operating pressure

Leaks

Tighten all fluid connections.

Insufficient solvent

Refill solvent reservoir.

Air in-line

Purge system.

Retention times too long

Pump flow rate set too low

Check flow setting and verify output.

Leaks

Tighten fluid connections.

Wrong columns

Verify length and diameter.

Retention times too short

Pump flow rate set too fast

Check flow rate setting and verify output.

Data collection device improperly connected Ensure data collection starts at inject signal.
Wrong columns

Verify length and diameter.

No peaks, poor recovery

Solvent flow too low

Verify pump pressure.

Injector blocked or leaking

Check for leaks or blockage.

Detector malfunctioning

Check wavelength, sensitivity, signal output cables.

Broad, smeared peaks, or peaks
appear as doublet

Rearrangement of packed bed or void in the
column packing

Check system performance with each column separately.
Reverse flow through affected column using 2 mL injections
of toluene or xylene at 2 mL/min. If not corrected after 10
injections, replace column.

Blockage or leak in injector sample loop

Check for leaks or excessive pressure drop across sample
loop. Back flush loop.

In-line or inlet filter is partially clogged

Reverse flow through column and inject 2 mL of
tetrahydrofuran, butylchloride, or a mixture of 1:1
cyclohexane:methylene chloride. Change the inlet filters.

Retention times shift more than
5%

Change in flow rate

Check flow rate.

Change in laboratory temperature

Stabilize laboratory temperature or column temperature.