beautypg.com

Care and use manual, B. band spreading minimization, D. measuring system volume – Waters Spherisorb Columns User Manual

Page 9: 36 μl

background image

[ Care and Use ManUal ]

Waters Spherisorb Columns

9

b. Band Spreading Minimization

Figure 6 shows the influence of tubing internal diameter on system
band spreading and peak shape. As can be seen, the larger tubing
diameter causes excessive peak broadening and lower sensitivity.

Diluted/Distorted Sample Band

0.005 inches

0.020 inches

0.040 inches

Figure 6: Effect of Connecting Tubing on System

c. Measuring System Bandspreading Volume and System
Variance

This test should be performed on an HPLC system with a single
wavelength UV detector (not a Photodiode Array [PDA]).

1. Disconnect column from system and replace with a zero dead

volume union.

2. Set flow rate to 1 mL/min.

3. Dilute a test mix in mobile phase to give a detector sensitivity of 0.5 -

1.0 AUFS (system start up test mix can be used which contains uracil,
ethyl and propyl parabens; Waters P/N WAT034544).

4. Inject 2 to 5 μL of this solution.

5. Measure the peak width at 4.4% of peak height (5-sigma method):

5-sigma Bandspreading (μL) = Peak Width (min) x Flow Rate (mL/min) x (1000 μL/1 mL)

System Variance (μL

2

) = (5-sigma bandspreading)

2

/ 25

System Volume

Figure 8: Determination of System Bandspreading Volume Using
5-Sigma Method

In a typical HPLC system, the bandspreading volume should be 100 μL
± 30 μL (or Variance of 400 μL

2

+/- 36 μL

2

). In a microbore (2.1 mm

i.d.) system, the bandspreading volume should be no greater than 20
to 40 μL (or variance no greater than 16 μL

2

to 64 μL

2

).

d. Measuring System Volume

System volume is important in scaling separations because it
creates an isocratic hold at the start of every run. This hold is often
several column volumes on a small scale, but a fraction of the
volume of a preparative column. Compensation for this volume must
be included in planning a scaling experiment to avoid distorting the
chromatography (Figure 9).

Figure 9: Determination of Gradient Delay Volume

Programmed time = 5 min

Flow Rate = 1.5 mL/min

5.69 min

-5.00 min

0.69 min

Time = 5.69 minutes

System Volume:

0.69 min x 1.5 mL/min = 1.04 mL

50%

0.70
0.65
0.60
0.55
0.50
0.45
0.40
0.35
0.30
0.25
0.20
0.15
0.10
0.05
0.00

AU

0

2

4

6

8

10

12

14

16

18

20 Min.