beautypg.com

Care and use manual, C. solvents, D. pressure – Waters XSelect HSS XP 2.5 µm Columns User Manual

Page 6: E. temperature

background image

XSelect HSS XP 2.5 µm Columns

6

[ CARE AND USE MANUAL ]

Table 5: Buffer Recommendations for
XSelect HSS XP 2.5 µm Columns

Additive or
Buffer

pKa

Buffer
Range

(±1 pH

unit)

Volatility

Used for

Mass

Spec?

Comments

TFA

0.3

Volatile

Yes

Ion pair additive, can suppress
MS signal. Used in the 0.01-
0.1% range.

Formic
Acid

3.75

Volatile

Yes

Maximum buffering obtained
when used with ammonium
formate salt. Used in 0.1-1.0%
range.

Acetic Acid

4.76

Volatile

Yes

Maximum buffering obtained
when used with ammonium
acetate salt. Used in 0.1-1.0%
range.

Formate
(HCOO-)

3.75

2.75 –

4.75

Volatile

Yes

Used in the 1-10 mM range.
Note: sodium or potassium salts
are not volatile.

Acetate
(CH

3

COO-)

4.76

3.76 –

5.76

Volatile

Yes

Used in the 1-10 mM range.
Note: sodium or potassium salts
are not volatile.

Phosphate 1

2.15

1.15 –

3.15

Non-

volatile

No

Traditional low-pH buffer, good
UV transparency.

Phosphate 2 7.2

6.20 –

8.20

Non-

volatile

No

Much shorter colum lifetimes
will be realized using phosphate
at pH 7.

c. Solvents

To maintain maximum column performance, use high quality HPLC
or MS grade solvents. Filter all aqueous buffers prior to use through
a 0.2 µm filter. Solvents containing suspended particulate materials
will generally clog the outside surface of the inlet of the column. This
may result in higher backpressure or distorted peak shape.

d. Pressure

XSelect HSS XP 2.5 µm Columns are compatible with HPLC, UHPLC
and UPLC pressures. Table 6 depicts the maximum operation pressure

Table 6: Maximum Operation Pressure

Column ID

Pressure Range

2.1 mm

18,000 psi [1034 bar]

3.0 mm

18,000 psi [1034 bar]

4.6 mm

9000 psi [620 bar]

e. Temperature

XSelect HSS XP 2.5 µm Columns can be used up at intermediate
temperatures to enhance selectivity, reduce solvent viscosity and
increase mass transfer rates.

Chemistry

Temperature Limit

XSelect HSS Cyano

45 °C

XSelect HSS PFP

45 °C

XSelect HSS T3

45 °C

XSelect HSS C

18

SB

45 °C

XSelect HSS C

18

45 °C

Note: Working in combinations of extreme pH, temperature and pressure may
result in reduced column lifetime.

IV. COLUMN CLEANING, REGENERATION AND STORAGE

a. Cleaning and Regeneration

Changes in peak shape, peak splitting, shouldering peaks, shifts
in retention, change in resolution or increasing backpressure may
indicate contamination of the column. Flush with a neat organic
solvent to remove the non-polar contaminant(s), taking care not to
precipitate any buffered mobile-phase components. If this flushing
procedure does not solve the problem, purge the column with the
following cleaning and regeneration procedures.

Use a cleaning routine that matches the properties of the samples,
stationary-phase type (reversed-phase, normal-phase or HILIC) and
will solubilize the suspected contaminate. Flush with 20 column
volumes of solvent at an intermediate temperature of 45 °C. Return
to the initial mobile-phase conditions by reversing the sequence.

If using a reversed-phase column, purge the column with a sequence
of progressively more non-polar solvents (i.e., water-to-methanol-to-
tetrahydrofuran-to-methylene chloride).

If using a HILIC column, purge the column with a
sequence of progressively more polar-organic solvents
(i.e., acetonitrile-to-acetonitrile/methanol-to-acetonitrile/
water-to-water).

If column performance has not improved after regeneration/
cleaning procedures, contact your local Waters representative
for additional support.