Care and use manual – Waters XSelect HSS XP 2.5 µm Columns User Manual
Page 10

XSelect HSS XP 2.5 µm Columns
10
[ CARE AND USE MANUAL ]
c. Recommended Flow Rates and Anticipated Backpressures for Reversed-Phase XSelect HSS XP 2.5 µm Columns
XP 2.5 µm, 2.1 mm ID Columns (40 °C)
Linear Velocity
3 mm/sec
4 mm/sec
5 mm/sec
6 mm/sec
Column Dimension
Flow Rate
[mL/min]
Backpressure
[psi]
Flow Rate
[mL/min]
Backpressure
[psi]
Flow Rate
[mL/min]
Backpressure
[psi]
Flow Rate
[mL/min]
Backpressure
[psi]
2.1 x 30 mm
0.45
1760
0.6
2350
0.75
2940
0.9
3520
2.1 x 50 mm
0.45
2640
0.6
3520
0.75
4400
0.9
5280
2.1 x 75 mm
0.45
3740
0.6
4980
0.75
6230
0.9
7470
2.1 x 100 mm
0.45
4830
0.6
6440
0.75
8055
0.9
9670
XP 2.5 µm, 3.0 mm ID Columns (40 °C)
Linear Velocity
3 mm/sec
4 mm/sec
5 mm/sec
6 mm/sec
Column Dimension
Flow Rate
[mL/min]
Backpressure
[psi]
Flow Rate
[mL/min]
Backpressure
[psi]
Flow Rate
[mL/min]
Backpressure
[psi]
Flow Rate
[mL/min]
Backpressure
[psi]
3.0 x 30 mm
0.9
2180
1.17
2840
1.53
3710
1.8
4360
3.0 x 50 mm
0.9
3040
1.17
3950
1.53
5170
1.8
6080
3.0 x 75 mm
0.9
4120
1.17
5350
1.53
7000
1.8
8230
3.0 x 100 mm
0.9
5190
1.17
6750
1.53
8825
1.8
10380
XP 2.5 µm, 4.6 mm ID Columns (40 °C)
Linear Velocity
3 mm/sec
4 mm/sec
5 mm/sec
6 mm/sec
Column Dimension
Flow Rate
[mL/min]
Backpressure
[psi]
Flow Rate
[mL/min]
Backpressure
[psi]
Flow Rate
[mL/min]
Backpressure
[psi]
Flow Rate
[mL/min]
Backpressure
[psi]
4.6 x 30 mm
2.1
3360
2.8
4480
3.5
5600
4.2
6720
4.6 x 50 mm
2.1
4210
2.8
5620
3.5
7020
4.2
8430
4.6 x 75 mm
2.1
5280
2.8
7040
3.5
8800
4.2
10560
4.6 x 100 mm
2.1
6350
2.8
8460
3.5
10580
4.2
12700