Care and use manual, E. measuring system dwell volume, F. column equilibration – Waters XSelect HSS XP 2.5 µm Columns User Manual

Page 3

background image

XSelect HSS XP 2.5 µm Columns

3

[ CARE AND USE MANUAL ]

System Volume

4.4 %h

5

Figure 2: Determination of system band spread volume using 5-Sigma Method.

Table 1: Expected System Band Spread Volumes

System

Band Spread

Alliance 2695 HPLC

29 µL

Vendor A HPLC

41 µL

Vendor B UHPLC (600 bar)

28 µL

Vendor C UHPLC

21 µL

Vendor D UHPLC

17 µL

ACQUITY UPLC

12 µL

ACQUITY UPLC H-Class

9 µL

ACQUITY UPLC I-Class (FTN)

7.5 µL

ACQUITY UPLC I-Class (FL)

5.5 µL

e. Measuring System Dwell Volume

Dwell volume is different than system band spreading. System dwell volume
is a measurement of the volume it takes for the initial gradient conditions to
reach the head of the column. This calculation is particularly useful when it is
necessary to transfer a method between different LC systems.

1. Disconnect the column from the system and replace with a

zero dead volume union.

2. Use acetonitrile as mobile phase A, and acetonitrile with

0.05 mg/mL uracil as mobile phase B.

3. Monitor UV at 254 nm.

4. Use the flow rate in the original method and the intended flow

rate on the target instrument.

5. Collect 100% A baseline for 5 minutes.

6. At 5 minutes, program a step to 100% B, and collect data for an

additional 5 minutes.

7. Measure absorbance difference between 100% A and 100% B.

8. Measure time at 50% of that absorbance difference.

9. Calculate time difference between start of step and 50% point.

10. Multiply time difference by flow rate to calculate system volume.

Programmed time = 5 minutes

50% Absorbance = 0.35852 AU
Time = 5.6953 minutes

100% Asymptotic

Total absorbance = 0.7164 AU

5.69
5.00
0.69 min

= Programmed Gradient

= Actual Gradient

System Volume
0.69 min x 1.5 mL/min = 1.04 mL

0.70

0.65

0.60

0.55

0.50

0.45

0.40

0.35

0.30

0.25

0.20

0.15

0.10

0.05

0.00

0.00

2.00

4.00

6.00

8.00

10.00

12.00

14.00

16.00

18.00

20.00

Figure 3: Measuring system band spread volume.

Table 2: Expected System Dwell Volumes

System

Dwell Volume

Alliance 2695 HPLC

900 µL

ACQUITY UPLC

120 µL

ACQUITY UPLC H-Class

350 µL

ACQUITY UPLC I-Class (FTN)

100 µL

ACQUITY UPLC I-Class (FL)

95 µL

f. Column Equilibration

XSelect HSS XP 2.5 µm Columns are shipped in 100% acetonitrile. It is
important to ensure mobile-phase compatibility before changing to a
different mobile-phase system. Equilibrate the column with a minimum
of 10 column volumes of the mobile phase to be used (refer to Table
3 for a list of column volumes). The column may be considered fully
equilibrated once a constant backpressure is achieved.