beautypg.com

Care and use manual – Waters XSelect HSS XP 2.5 µm Columns User Manual

Page 5

background image

XSelect HSS XP 2.5 µm Columns

5

[ CARE AND USE MANUAL ]

Note: VanGuard Pre-Columns are shipped with a collet and ferrule that are NOT
pre-swaged. This enables the end user to mate the VanGuard Pre-Column to
a specific XP 2.5 µm Column and ensures void-free and leak-free connections.
Care must be taken when removing the O-ring that holds these two pieces
on the pre-column tubing.

XP 2.5 µm Column

VanGuard Pre-Column

Place wrench here

Ferrule

Collet

Place wrench here

Flow

Figure 4: Installing a VanGuard Pre-Column.

VanGuard Pre-Column Installation Instructions

1. Remove the VanGuard Pre-column from its box and shipping

tube and remove the plastic plug.

2. Orient the pre-column so that the male end is facing up and

carefully remove the black O-ring that holds the collet and
ferrule in place during shipment (collet and ferrule are not
permanently attached).

3. Orient the XP 2.5 µm Column perpendicular to the work surface

so that the column inlet is on the bottom.

4. From below, insert the VanGuard Pre-Column into the column

inlet; turn the assembled column and pre-column 180° so that
the pre-column is now on top.

5. Tighten with two 5/16” wrenches placed onto the XP Column

flats and VanGuard Pre-Column hex nut (male end) as shown in
Figure 4.

6. While keeping pressure on the VanGuard Pre-Column against the

XP 2.5 µm Column, tighten turn to set the collet and ferrule.

7. Check that the ferrule is set by loosening the connection and

inspecting the ferrule depth.

8. Reattach the pre-column to the XP 2.5 µm Column, apply flow

and inspect for leaks.

III. COLUMN USE

To ensure the continued high performance of XSelect HSS XP 2.5 µm
Columns, follow these guidelines:

a. Sample Preparation

1. Sample impurities and/or particulates often contribute to column

contamination. One option to avoid column contamination is to
use Waters Oasis

®

or Sep-Pak

®

Solid-Phase Extraction (SPE)

devices. To select the appropriate sorbent for a specific sample
type, visit www.waters.com/sampleprep

2. It is preferable to prepare the sample in the initial mobile-

phase conditions or a weaker solvent for the best peak shape
and sensitivity.

3. If the sample is not prepared in the mobile phase, ensure that

the sample, solvent and mobile phases are miscible in order to
avoid sample and/or buffer precipitation.

4. Filter sample with a 0.2 µm membrane to remove particulates.

If the sample is dissolved in a solvent that contains an organic
modifier (e.g., acetonitrile, methanol, etc.) ensure that the
membrane/filter material is compatible with the solvents
in use. Alternatively, centrifuge the sample for 20 minutes at
8000 rpm, followed by the transfer of the supernatant to an
appropriate vial could be considered.

b. pH Range


Table 4: Recommended pH Range

Chemistry

pH Range

XSelect HSS Cyano

2 - 8

XSelect HSS PFP

2 - 8

XSelect HSS T3

2 - 8

XSelect HSS C

18

SB

2 - 8

XSelect HSS C

18

1 - 8

Column lifetime will vary depending on the combination of temperature,
mobile-phase pH and type of buffer/additive used. Table 5 lists the
recommended buffers and additives for XSelect HSS XP 2.5 µm Columns.

Note: Working in combinations of extreme pH, temperature and pressure may result in
reduced column lifetime.