Care and use manual – Waters ACQUITY UPLC CSH130 C18 1.7 μm and XP 2.5 μm Columns User Manual
Page 8

[ CARE AND USE MANUAL ]
ACQUITY UPLC CSH130 C
18
Columns
8
VI. ADDITIONAL INFORMATIONA.
Tips for maximizing ACQUITY UPLC CSH130 C
18
Column Lifetimes
1. To maximize ACQUITY UPLC CSH130 C
18
column lifetime, pay close
attention to:
• Water quality (including water purification system)
• Solvent quality
• Mobile-phase preparation, storage and age
• Sample, buffer and mobile-phase solubilities
• Sample quality and preparation.
2. When problems arise, often only one improper practice must be changed.
3. Always remember to:
• Use in-line filter unit or, preferably, a VanGuard Pre-column.
	 •	 Discourage	bacterial	growth	by	minimizing	the	use	of	100% 
 aqueous mobile phases where possible.
	 •	 Change	aqueous	mobile	phase	every	24-48	hours	(if	100% 
 aqueous mobile phase use is required). 
	 •	 Discard	old	100%	aqueous	mobile	phases	every	24-48	hours 
 to discourage bacterial growth.
	 •	 Add	5%-10%	organic	modifier	to	mobile	phase	A	and	adjust	 
 gradient profile. 
• Filter aqueous portions of mobile phase through 0.2 µm filter.
	 •	 Maintain	your	water	purification	system	so	that	it	 is	in	good 
 working order.
	 •	 Only	use	ultra	pure	water	(18	megohm-cm)	water		and	highest	quality	 
 solvents possible. HPLC grade water is not UPLC grade water.
• Consider sample preparation (e.g., solid-phase extraction, filtration, etc).
4. Avoid (where possible):
• 100% aqueous mobile phases (if possible).
• HPLC-grade bottled water.
• “Topping off” or adding “new” mobile phase to “old” mobile phase.
	 •	 Old	aqueous	mobile	phases.	Remember	to	rinse	bottles	thoroughly	and 
 prepare fresh every 24-48 hours.
	 •	 Using	phosphate	salt	buffer	in	combination	with	high	ACN	concentrations	 
 (e.g., > 70%) due to precipitation.
5. Don’t: assume a “bad” column is the culprit when high back pressure or split 
 peaks are observed. Investigate cause of column failure: 
• Backpressure
• Mobile phase(s), bacteria, precipitation and/or samples
• Peak splitting
• Sample quality
• Injection solvent strength.
6. Remember: UPLC flow rates are often much lower and, therefore, 
 mobile phases last much longer (only prepare what you need or store 
 excess refrigerated).
7. Mobile-phase related questions to ask:
	 •	 Am	I	using	100%	aqueous	mobile	phases?	Am	I	able	to	add	 
	 	 a	small	amount	of	organic	modifier	to	my	mobile-phase	A?
• Do I filter my aqueous mobile phases through 0.2 µm filters?
• How old is my mobile phase? Do I label the bottle with preparation date?
• Do I “top off” or do I prepare fresh mobile phases every 24-48 hours?
	 •	 What	is	the	quality	of	my	water?	Has	the	quality	recently	changed?	How	 
	 	 is	my	water	purification	system	working?	When	was	it	last	serviced?
	 •	 Am	I	working	with	a		pH	7	phosphate	buffer	(which	is	VERY	susceptible 
	 	 to	bacterial	growth)?
8. Sample-related questions to ask:
	 •	 If	I	inject	neat	standards	prepared	in	mobile	phase	do	I	observe 
	 	 these	problems?
	 •	 If	I	prepare	my	standards	in	water	and	prepare	them	like	samples 
	 	 (e.g.,	SPE,	filtration,	etc.)	do	I	still	observe	these	problems?
• Has the quality of my samples changed over time?
