beautypg.com

Care and use manual – Waters ACQUITY UPLC CSH130 C18 1.7 μm and XP 2.5 μm Columns User Manual

Page 7

background image

[ CARE AND USE MANUAL ]

ACQUITY UPLC CSH130 C

18

Columns

7

1.20
1.10

1.00
0.90
0.80

0.70

0.60
0.50

0.40

0.30
0.20

0.10

0.00

2

3

4

5

6

7

8

9

10

11

12 Min.

V. REPRESENTATIVE TEST CHROMATOGRAPH AND CONDITIONS FOR SEPARATION OF A CYTOCHROME C TRYPTIC DIGEST

Peak Identification

T1

N-AcGDVEK

T13-T14

KYIPGTK

T14

YIPGTK

T4

IFVQK

T9-T10

KTGQAPGFSYTDANK

T10

TGQAPGFSYTDANK

T8

TGPNLHGLFGR

T15

MIFAGIK

T19C

EDLIAY

T19

EDLIAYLK

T12-T13

GITWGEETLMEYLENPKK

T12

GITWGEETLMEYLENPK

T5

CAQCHTVEK (heme attached)

Retention Time (min)

Result

T19C

6.71

Retention Time Difference (min)

Result

T4 and T14

0.30

T10 and T9-T10

0.77

T19 and T19C

0.47

T12 and T12-T13

0.28

Chromatographic Conditions  

Column:

2.1 mm x 50 mm

Mobile Phase A: 0.1% formic acid in 100% water

Mobile Phase B: 0.085% formic acid in 75% acetonitrile reconstituted

w/ 200 µL of MPA, 5.0 µL injection

Sample:

Tryptic digest of bovine cytochrome c (P/N 186006371)

Flow rate:

0.2 mL/min

Gradient:

1–21% B in 5 min; 21–31.4% B in 2.5 min; 31.4–95% B

in 2.5 min; 95% B for 1.5 min; 1%B for 2.5 min

Temperature:

40 °C Detection: UV 214 nm