Modifying the instrument configuration (continued)

Standard setting is "NO".

• To change the setting, press the PLUS key.

When answering "YES", programs no. 6 - 9 are

set to standard values and programs no. 1 - 5 are

deleted.

• For the next menu item, press the CURSOR

button.

The next standard setting is displayed.

SystemReset? =

Reset a number of settings

to standard value

Standard setting is "NO".

• To change the setting, press the PLUS key.

When answering "YES", the system clock is set

at 0:00 and the maximum heating time for the

wax baths is set at 8 hours. No other settings

are altered.

Leaving the configuration menu
• To leave the configuration menu, press PROG.

Programs no. 6 - 9 come with a number

of factory preset standard values; how-

ever these may be altered by the user.

A system reset will set real time to 0.00

and maximum heating time for the wax

baths to 8 hours. The "Sound Signal

Level" setting and the number of wax

baths selected in the standard configu-

ration menu will remain unaltered by a

system reset.