Appendix – Leica Biosystems TP1020 User Manual
Page 92
92
Instructions for Use V 2.9 RevC, English – 12/2013
15. Appendix
Modifying the instrument configuration (continued)
Standard setting is "NO".
• To change the setting, press the PLUS key.
When answering "YES", programs no. 6 - 9 are
set to standard values and programs no. 1 - 5 are
deleted.
• For the next menu item, press the CURSOR
button.
The next standard setting is displayed.
SystemReset? =
Reset a number of settings
to standard value
Standard setting is "NO".
• To change the setting, press the PLUS key.
When answering "YES", the system clock is set
at 0:00 and the maximum heating time for the
wax baths is set at 8 hours. No other settings
are altered.
Leaving the configuration menu
• To leave the configuration menu, press PROG.
Programs no. 6 - 9 come with a number
of factory preset standard values; how-
ever these may be altered by the user.
A system reset will set real time to 0.00
and maximum heating time for the wax
baths to 8 hours. The "Sound Signal
Level" setting and the number of wax
baths selected in the standard configu-
ration menu will remain unaltered by a
system reset.