beautypg.com

2 error codes, Warning and error codes – troubleshooting – Leica Biosystems TP1020 User Manual

Page 70

background image

70

Instructions for Use V 2.9 RevC, English – 12/2013

10. Warning and Error Codes – Troubleshooting

10.2 Error codes

To indicate a series of malfunctions, the error

codes E:01 to E:13 are displayed.

Three consecutive and very short sound signals

that are repeated every 10 seconds reinforce the

visual error code.

• To quit the error codes, press any key.

List of error codes

Code

Possible root cause

Troubleshooting

Error code "E:01"

Cache memory defective

Call Technical Service

Error code "E:03" -

"E:04" - "E:05"

Drive malfunction

- Switch off the main switch and

next switch it back on.

Try again.

- Call Technical Service

Error code "E:07"

Clock component defective

Call Technical Service

Error code "E:08"

(Wax bath 1)

Error code "E:09"

(Wax bath 2)

Error code "E:10"

(Wax bath 3)

- Wax bath not connected.

- Wax bath possibly connected to

wrong socket.

- Excess temperature switch-off

mechanism has responded.

- Wax bath defective.

- Control unit defective.

- Connect wax bath.

- Check if wax bath is connected to

the appropriate socket. Correct if

necessary.

- Disconnect the wax bath connec-

tion plug from the socket at the

rear of the instrument.

Let wax bath cool down.

Reconnect wax bath

If the yellow pilot lamp lights, use

of the wax bath may be continued.

- Exchange wax bath.

- Call Technical Service

On the next page you will find a com-

plete list of all error messages, their

meaning and how to troubleshoot them.