Instructions for Use V 2.9 RevC, English – 12/2013

Wax baths

Maintenance

Activated carbon filter (optional accessory)

2. Safety

• Only authorized and qualified service personnel may access the internal components of the

instrument for service and repair!

• Before replacing the fuses, turn off the power switch and unplug the instrument!

• Burnt-out fuses may only be replaced by fuses of the same type and specification. For appro-

priate brands and specifications, see

Chapter 3 "Technical Data".

• If the excess temperature repeatedly - please call Leica Technical Service. DO NOT continue

to use the wax bath.

• Check if the wax bath works trouble-free. Use of defective wax baths has to be discontinued

for safety reasons!

• Empty the wax baths completely before the paraffin cools! When cooling down, the paraffin

contracts, which can cause damage to the wax bath.

• Warning! Fire hazard! If a filter becomes saturated with solvent, there is a potential fire risk!

It is important that the activated carbon filters are changed at factory recommended inter-

vals.