beautypg.com

Warning and error codes – troubleshooting – Leica Biosystems TP1020 User Manual

Page 69

background image

69

Leica TP1020

10. Warning and Error Codes – Troubleshooting

+

• To override the warning codes W:04, W:05 and

W:06 press KEY and START simultaneously to

start processing.

Warning code (continued)

W:01 to W:03 block transportation of the tissue basket into or out of the wax bath of station 10,

11 or 12. However, if you are certain that the paraffin is actually liquefied, you can manually im-

merse the tissue basket into or remove it from the wax bath.

Code

Possible root cause

Troubleshooting

Warning code

W:01, W:02, W:03

in manual pro-

cessing mode

Paraffin in wax bath stations 10/11/12

still solid.

- To quit the warning code, press any

key.

- Check if the paraffin is really molten.

- If you find that the paraffin is molten,

pause the automatic processing

cycle and - in the manual process-

ing mode - by pressing KEY and

the corresponding ARROW button

simultaneously, lower the basket

into the wax bath or lift it out of the

wax bath.

Warning code

" W : 0 4 " " W : 0 5 "

"W:06"

when starting an

automatic pro-

cessing cycle.

Total duration of program from start un-

til the basket reaches the first wax bath

is less than 8 hours. Paraffin in stations

10/11/12 possibly still solid upon arrival

of tissue basket.

- Check if there is sufficient time for

the paraffin to liquefy completely

until the basket arrives at the wax

bath.

- If this is not ensured, fill the wax bath

with molten paraffin.

- To override / quit the warning code,

press KEY and START simultane-

ously.