Instructions for Use V 2.9 RevC, English – 12/2013
10. Warning and Error Codes – Troubleshooting
10.5 "ABORT" message
"ABORT" is displayed after quitting an error code
emitted due to instrument malfunction. When
confirming such an error message processing
is aborted.
At this point processing can be restarted from the
beginning. To skip the stations that had already
been covered during the previous aborted run
proceed as follows:
• Press START.
• To start the currently displayed program, press
START again.
This will cause the instrument to step by step
move the basket to the programmed starting sta-
tion. Before reaching that position:
• Press PAUSE/CONT to switch to the manual
processing mode.
• Press CIRCLE ARROW to move the basket to
the station where processing was aborted.
• Press ARROW DOWN to lower the basket.
• To leave the manual processing mode, press
PAUSE/CONT again.
Processing will then continue where it had previ-
ously been aborted.