Leica TP1020

• Activate the programming mode.

• Select a program.

To verify and/or change the selected parameters,

• Move the cursor to the station number entry position (ARROW RIGHT /

LEFT = cursor buttons)

• Modify the station number pressing PLUS or MINUS and thus edit, and

if necessary alter, step by step all station parameters.

9. Operating the Instrument in the Automatic Processing Mode

9.2 Editing and modifying programs while a processing cycle is in progress

• Press PROG to quit programming.

9.3 Display end of processing time

While a processing cycle is in progress, it is possible to display the

expected end of run time.

• To display end of run time, press CLOCK.

Programs can be edited and changed while a processing cycle is in

progress. All programs can be edited and changed with the excep-

tion of the program currently in progress. A program in progress can

be displayed but not changed.

All modifications are saved immediately. The program currently in

progress cannot be changed!