beautypg.com

5 resuming a paused process, 6 stopping or terminating a process, Resuming a paused process – Leica Biosystems TP1020 User Manual

Page 65: Stopping or terminating a process

background image

65

Leica TP1020

9. Operating the Instrument in the Automatic Processing Mode

9.5 Resuming a paused process

9.6 Stopping or terminating a process

• To reassume processing press PAUSE/CONT

again.

Pressing this key will lower the basket into the

station. Processing continues as programmed.
While a run is paused, the infiltration time count-

down in the actual station will also be interrupt-

ed. When processing is resumed, the remaining

infiltration time in the station will be carried out

without interruption. Thus, the end of run time of

a process will be postponed by the length of time

the process was paused.
The display reading will automatically be

updated, showing the new end of run time.
If there is an emergency, an automatic process-

ing cycle can be stopped immediately and - in a

subsequent step - entirely aborted.

• To stop a processing cycle in progress press

STOP.

The carousel up-and-down movement stops im-

mediately.
The display reads "STOP?".
With this the instrument is asking you whether

you really want to abort the processing cycle

currently in progress?

• To resume processing, press START.
Processing will continue as programmed.