4 wax bath, Instrument components and specifications – Leica Biosystems TP1020 User Manual

Page 25

background image

25

Leica TP1020

3. Instrument Components and Specifications

1

Fig. 9

3

2

4

5

Holding capacity:

1.8 l

Nominal supply voltages:

230V AC

50 - 60 Hz

Nominal capacity (heating):

150 VA

Temperature setting range:

45 - 65 °C ±3 K

Excess temperature switch-off

mechanism:

at 85 °C ±5 K

3.6.4 Wax bath

If the actual temperature rises above the range of

normal working temperature, an excess tempera-

ture switch-off mechanism responds. The wax

bath heating is disconnected. The yellow signal

lamp is extinguished. Use of the wax bath can be

resumed only after a cooling down period. For

cooling down, disconnect the wax bath connector

plug from the socket at the rear of the instrument,

or switch off the main switch.

The wax baths are equipped with heating. While

the heating is functioning, the yellow pilot lamp

lights (1).
The handle (2) is insulated. The outside wall

(3) is likewise thermally insulated. However, it

becomes hot to the touch in heating operation.

Despite this, burn hazard exists when touching it.

The inside of the interior container (4) is coated

with Teflon. The interior container gets very hot.

It has minimum and maximum level indicator

marks (5).

Caution! The interior container (4) of

the wax bath becomes very hot when

the heating function is activated! Do

not touch the gray upper rim of the con-

tainers with your hands! Risk of injury!

If the excess temperature switch-off

mechanism reacts several times, call

Technical Service without delay. DO

NOT continue to use the wax bath.

Empty the wax baths completely be-

fore the paraffin cools! When cooling

down, the paraffin contracts, which

can cause damage to the wax bath.