15.1 Modifying the instrument configuration
+
The instrument has a factory-set standard con-
figuration which can be modified by the user if
desired.
Modifications can be done in the configuration
menu, which consists of 4 different menu items:
- Signal level 1 - 4
- 3. Wax bath?
- ProgPreset?
- SystemReset?
Editing the configuration menu
• Switch the instrument off at the main power
switch.
• Turn the power switch back on. While the
display still reads "TP1020 V x.xx" press PROG
for a short while.
The first configuration menu item is displayed.
Modifying the configuration
Signal level = Setting of the desired type and
frequency of sound signal
Level 1 = sound signal in case of power failure
and malfunctions.
Level 2 = level 1 + sound signal in case of inter-
rupt ion, warning code and end of
program.
Level 3 = level 2 + sound signal when confirm-
ing warning codes and error codes.
Level 4 = level 3 and sound signal every time a
button is pressed.
15. Appendix
For this menu item four different settings
can be selected. Setting 4 is factory
preset.
Setting 1 is the lowest level. The next
higher level always contains all the
parameters of the next preceding level
plus the additional signals as listed on
this page.