2 setting-up / editing programs, 1 selecting programming mode, 2 selecting the program – Leica Biosystems TP1020 User Manual
Page 45: 3 selecting the number of baskets, Setting-up / editing programs, Programming the instrument
![background image](/manuals/565432/45/background.png)
45
Leica TP1020
6. Programming the Instrument
6.2 Setting-up / editing programs
6.2.1 Selecting programming mode
6.2.2 Selecting the program
6.2.3 Selecting the number of baskets
• To select the programming mode, push the
PROG key.
The following parameters are displayed:
- Number of baskets or
,
- Number of processing station 1 - 12,
- Vacuum "ON" or "OFF",
- Tissue infiltration time in the displayed sta-
tion "0 h 50 min",
- Program number
"P1 - P9",
- Programming mode "PROG".
The cursor blinks at the program number.
• Select the desired program number by
pressing PLUS or MINUS.
• Push the ARROW LEFT button to move the
cursor to the upper line.
The cursor blinks next to the first basket symbol.
• To change the number of baskets from 1 to 2,
press PLUS.
• To change the number of baskets from 2 to 1,
press MINUS.
With this, the infiltration for the tissue baskets will