Leica TP1020
10. Warning and Error Codes – Troubleshooting
10.4 Measures to remove specimen material in case of prolonged power failure
Fig. 23
2
1
• Switch the instrument off at the main power
switch.
• Remove the crank handle (1) from the clip and
insert it into the access port (2) on the left side
of the instrument.
Removing the tissue basket
• Lift the carousel rotating the
crank handle in the direction of
the arrow until the lower end of
the basket is above the rim of the
processing station.
• Hang on to the crank handle to
keep it securely in its position
and remove the tissue basket.
In case of a prolonged power failure
the tissue can be removed from the
instrument manually and also be re-
located to the next station by entirely
manual operation.
- SM2010 R (46 pages)
- CM305 S (66 pages)
- CE Knife holder (14 pages)
- CM1950 (78 pages)
- VT1200/VT1200 S (62 pages)
- VT1000 S (42 pages)
- HI1220 (18 pages)
- HI1210 (18 pages)
- CM1850 UV (56 pages)
- CM1850 UV (54 pages)
- CM1510-S (56 pages)
- CM3600 XP (114 pages)
- CM1100 (36 pages)
- CM1860/CM1860 UV (98 pages)
- Premium blade holder (14 pages)
- CM1520 (92 pages)
- VT1000 A (49 pages)
- VT1000 P (48 pages)
- SP2600 (51 pages)
- SM2500 (76 pages)
- SM2400 (8 pages)
- SM2000R (22 pages)
- EG1160 (32 pages)
- EG1150 C (20 pages)
- EG1150 H (42 pages)
- EG1130 (2 pages)
- EG1120 (2 pages)
- EG F (18 pages)
- ASP300 (70 pages)
- PELORIS_PELORIS II (160 pages)
- ASP200 S (88 pages)
- ASP300 S (88 pages)
- ASP6025 (178 pages)
- LN22 (44 pages)
- RM CoolClamp (20 pages)
- RM2265 (88 pages)
- RM2255 (82 pages)
- RM2245 (78 pages)
- RM2235 (78 pages)
- RM2125 RTS (60 pages)
- SP9000 (30 pages)
- SP1600 (10 pages)
- ST4040 (50 pages)
- ST5020 (88 pages)