3 replacing the fuses, Maintenance – Leica Biosystems TP1020 User Manual

Page 81

background image

81

Leica TP1020

12.3 Replacing the fuses

12. Maintenance

Fig. 25

7

3

1

6

The fuses are located in the voltage selector

shell (3).

• Remove the shell (3) as described in Chap-

ter 4.3.2 "Adjusting the voltage selector".

• Remove the fuses (7).

• Insert replacement fuses of the same type.

The currently selected voltage can be seen in

the small window (1) in the voltage selector shell.

• Insert the voltage selector shell together with

the fuses back into the receptacle in the instru-

ment and press lightly until it locks.

• Check if the setting that shows in the window

(1) corresponds to the nominal voltage in your

laboratory.

Turn the instrument off and disconnect the power plug before replacing the fuses! Burnt-out

fuses may only be replaced by fuses of the same type and specification.

See Chapter 3 "Technical Data"

.