3 replacing the fuses, Maintenance – Leica Biosystems TP1020 User Manual
Page 81
81
Leica TP1020
12.3 Replacing the fuses
12. Maintenance
Fig. 25
7
3
1
6
The fuses are located in the voltage selector
shell (3).
• Remove the shell (3) as described in Chap-
ter 4.3.2 "Adjusting the voltage selector".
• Remove the fuses (7).
• Insert replacement fuses of the same type.
The currently selected voltage can be seen in
the small window (1) in the voltage selector shell.
• Insert the voltage selector shell together with
the fuses back into the receptacle in the instru-
ment and press lightly until it locks.
• Check if the setting that shows in the window
(1) corresponds to the nominal voltage in your
laboratory.
Turn the instrument off and disconnect the power plug before replacing the fuses! Burnt-out
fuses may only be replaced by fuses of the same type and specification.
See Chapter 3 "Technical Data"
.
- SM2010 R (46 pages)
- CM305 S (66 pages)
- CE Knife holder (14 pages)
- CM1950 (78 pages)
- VT1200/VT1200 S (62 pages)
- VT1000 S (42 pages)
- HI1220 (18 pages)
- HI1210 (18 pages)
- CM1850 UV (56 pages)
- CM1850 UV (54 pages)
- CM1510-S (56 pages)
- CM3600 XP (114 pages)
- CM1100 (36 pages)
- CM1860/CM1860 UV (98 pages)
- Premium blade holder (14 pages)
- CM1520 (92 pages)
- VT1000 A (49 pages)
- VT1000 P (48 pages)
- SP2600 (51 pages)
- SM2500 (76 pages)
- SM2400 (8 pages)
- SM2000R (22 pages)
- EG1160 (32 pages)
- EG1150 C (20 pages)
- EG1150 H (42 pages)
- EG1130 (2 pages)
- EG1120 (2 pages)
- EG F (18 pages)
- ASP300 (70 pages)
- PELORIS_PELORIS II (160 pages)
- ASP200 S (88 pages)
- ASP300 S (88 pages)
- ASP6025 (178 pages)
- LN22 (44 pages)
- RM CoolClamp (20 pages)
- RM2265 (88 pages)
- RM2255 (82 pages)
- RM2245 (78 pages)
- RM2235 (78 pages)
- RM2125 RTS (60 pages)
- SP9000 (30 pages)
- SP1600 (10 pages)
- ST4040 (50 pages)
- ST5020 (88 pages)