Leica Biosystems TP1020 User Manual
Page 61

61
Leica TP1020
9. Operating the Instrument in the Automatic Processing Mode
Delayed start (continued)
Edit and change starting time (when using the delayed starting function)
+
• Press CLOCK again.
• Check the selected starting time and if neces-
sary change it as appropriate to achieve the
desired end or run time.
• To activate the delay function press START.
• To override the warning codes and start pro-
cessing, press KEY and START simultaneously.
The program will be started at the selected delay
time.
• To quit the display indication, press CLOCK
again.
The warning codes W:04 - W:06 (see Chapter 7) are displayed for programs of less than 8 hours
duration from start until the basket reaches the first wax bath. Check if the available amount of
time is sufficient for the paraffin to liquefy completely. If this is not ensured, fill the wax baths
with liquid paraffin.
- SM2010 R (46 pages)
- CM305 S (66 pages)
- CE Knife holder (14 pages)
- CM1950 (78 pages)
- VT1200/VT1200 S (62 pages)
- VT1000 S (42 pages)
- HI1220 (18 pages)
- HI1210 (18 pages)
- CM1850 UV (56 pages)
- CM1850 UV (54 pages)
- CM1510-S (56 pages)
- CM3600 XP (114 pages)
- CM1100 (36 pages)
- CM1860/CM1860 UV (98 pages)
- Premium blade holder (14 pages)
- CM1520 (92 pages)
- VT1000 A (49 pages)
- VT1000 P (48 pages)
- SP2600 (51 pages)
- SM2500 (76 pages)
- SM2400 (8 pages)
- SM2000R (22 pages)
- EG1160 (32 pages)
- EG1150 C (20 pages)
- EG1150 H (42 pages)
- EG1130 (2 pages)
- EG1120 (2 pages)
- EG F (18 pages)
- ASP300 (70 pages)
- PELORIS_PELORIS II (160 pages)
- ASP200 S (88 pages)
- ASP300 S (88 pages)
- ASP6025 (178 pages)
- LN22 (44 pages)
- RM CoolClamp (20 pages)
- RM2265 (88 pages)
- RM2255 (82 pages)
- RM2245 (78 pages)
- RM2235 (78 pages)
- RM2125 RTS (60 pages)
- SP9000 (30 pages)
- SP1600 (10 pages)
- ST4040 (50 pages)
- ST5020 (88 pages)