3 replacing reagent – remote fill and drain, Replacing reagent – remote fill and drain – Leica Biosystems PELORIS_PELORIS II User Manual
Page 91

Reagent Setup
Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011
91
5.4.3 Replacing Reagent – Remote Fill and Drain
You can drain and then refill reagent bottles without removing the bottles from the instrument. The
process drains each bottle to a retort then drains the retort to the remote fill/drain line. The reverse
procedure is used for the fill.
If you routinely use remote drain and fill do not forget to check if bottles need cleaning, once a
week.
Before you begin a remote fill or drain ensure there is a retort available:
the retort must not have a protocol loaded or running;
The retort must be clean or empty;
The retort residue (if any) must be compatible with the reagent in the bottle(s).
Follow the instructions below to replace reagent in a bottle.
Drain
1. Connect the remote fill/drain hose and place the end in a suitable container (see 5.4.2 Remote
2. From the Remote fill/drain screen (Reagents menu, Remote fill/drain), select:
The retort to use
The Remote icon
The bottle(s) to drain (multiple bottles must all have reagent in the same group).
Figure 54. Remote fill/drain setup for draining bottles
3. Tap Drain from bottle(s) to begin the drain.
Select the retort to use
Select the bottle or bottles to
drain
Select Remote icon
When bottle, retort and
Remote are selected tap
Drain from bottle(s) to
start the drain