Xylene 8 hour – Leica Biosystems PELORIS_PELORIS II User Manual
Page 129

Reference
Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011
129
Xylene 6 Hour
Not a pre-defined protocol
Xylene 8 Hour
Pre-defined protocol “Factory 8hr Xylene Standard“
Carryover setting in pre-defined protocol: 25
Step Reagent type
Reagent
group
Time
(min)
Temp
(°C)
P/V
Stirrer
Drip time
(s)
1
Formalin
Fixatives
15
45
Ambient
Medium
10
2
Ethanol
Dehydrants
15
45
Ambient
Medium
10
3
Ethanol
Dehydrants
15
45
Ambient
Medium
10
4
Ethanol
Dehydrants
15
45
Ambient
Medium
10
5
Ethanol
Dehydrants
15
45
Ambient
Medium
10
6
Ethanol
Dehydrants
30
45
Ambient
Medium
10
7
Ethanol
Dehydrants
45
45
Ambient
Medium
10
8
Xylene
Clearers
20
45
Ambient
Medium
10
9
Xylene
Clearers
20
45
Ambient
Medium
10
10
Xylene
Clearers
45
45
Ambient
Medium
10
11
Paraffin wax
Wax
30
65
Vacuum
Medium
10
12
Paraffin wax
Wax
30
65
Vacuum
Medium
10
13
Paraffin wax
Wax
45
65
Vacuum
Medium
10
Processing time
6:08:00
Step Reagent type
Reagent
group
Time
(min)
Temp
(°C)
P/V
Stirrer
Drip time
(s)
1
Formalin
Fixatives
20
45
Ambient
Medium
10
2
Ethanol
Dehydrants
20
45
Ambient
Medium
10
3
Ethanol
Dehydrants
20
45
Ambient
Medium
10
4
Ethanol
Dehydrants
20
45
Ambient
Medium
10
5
Ethanol
Dehydrants
20
45
Ambient
Medium
10
6
Ethanol
Dehydrants
40
45
Ambient
Medium
10
7
Ethanol
Dehydrants
60
45
Ambient
Medium
10
8
Xylene
Clearers
30
45
Ambient
Medium
10
9
Xylene
Clearers
30
45
Ambient
Medium
10
10
Xylene
Clearers
60
45
Ambient
Medium
10
11
Paraffin wax
Wax
40
65
Vacuum
Medium
10
12
Paraffin wax
Wax
40
65
Vacuum
Medium
10
13
Paraffin wax
Wax
60
65
Vacuum
Medium
10
Processing time
8:08:00