2 remote fill/drain connections, Remote fill/drain connections – Leica Biosystems PELORIS_PELORIS II User Manual
Page 89

Reagent Setup
Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011
89
Figure 50 shows the Remote fill/drain screen with the main features described:
Figure 50. Remote fill/drain screen
5.4.2 Remote Fill/Drain Connections
The remote fill/drain line and the wax drain outlet sit above the carbon filter in the reagent cabinet
(see Figure 52). A protective flap covers the outlets. The wax waste line is heated to ensure that
the wax does not solidify during the drain.
Before draining a wax station, open the upper reagent cabinet door, slide back the fill/drain flap
then fit the wax waste hose to the wax waste line (right connection in Figure 52).
Ensure the wax waste hose drains into a suitable container.
Ensure the hose remains out of the drained wax so that wax can’t solidify on the end of hose.
Before filling or draining reagent bottles, connect the remote fill/drain hose (see Figure 53) to
the remote fill/drain line (left connection in Figure 53). The hose has a push-fit coupling that
ensures a secure connection to the line.
Fill to/Drain from bottles, Drain to waste, or
Fill/Drain retort:
start fill or drain when appropriate station, remote
source and retort are selected
Abandon:
stop the drain or fill
Wax waste:
select with a wax
chamber to drain the
chamber
Remote:
select with a retort and
bottle to fill or drain
the bottle
Retort A:
scheduled fill/drain
operations for retort A
Retort B:
scheduled fill/drain
operations for retort A
Wax System:
scheduled fill/drain
operations for the wax
chambers
Figure 51. Remote fill/drain flap closed
Figure 52. Remote fill/drain flap open with reagent
line (left) and wax waste line (right).