8.1 Reagent Threshold Guidelines
The tables in this section list the recommended thresholds for commonly used reagents. There are
separate tables for xylene and xylene-free processing:
Xylene Processing
For the best results, xylene-processing reagent changes should be based on concentration
thresholds, while cleaning reagent thresholds should be based on cycles.
Thresholds for xylene-processing reagents, including cleaning reagents, using nongraded alcohol
are:
Xylene-free Processing
In general, xylene-free processing reagent changes should be based on concentration thresholds,
and cleaning reagent changes based on cycles.
* This differs from the recommendation for cleaning ethanol in xylene processing because the cleaning solvent
in xylene-free protocols is less efficient than the cleaning solvent in xylene protocols.
Reagent change
thresholds
Final reagent
thresholds
Max. temperatures
C
Type
Conc. (%) Cassettes
or cycles
Conc. (%) Cassettes
or cycles
Ambient Vacuum Safe
Formalin
98.0
1500 cass. N/A
N/A
60
60
45
Ethanol
51.0
N/A
98.0
1500 cass. 78
51
45
Xylene
68.0
N/A
95.0
1500 cass. 138
99
45
Wax
80.0
4500 cass. 95.0
1500 cass. 100
100
77
Cleaning xylene
88.0
10 cycles N/A
N/A
138
99
45
Cleaning ethanol
88.0
10 cycles N/A
N/A
78
51
45
Reagent change
thresholds
Final reagent
thresholds
Max. temperatures
C
Type
Conc. (%) Cassettes
or cycles
Conc. (%) Cassettes
or cycles
Ambient Vacuum Safe
Formalin
98.0
1500 cass. N/A
N/A
60
60
45
85% ethanol
50.0
N/A
N/A
N/A
87
55
45
80/20 ethanol/IPA
81.0
N/A
N/A
N/A
78
51
45
IPA
90.0
4500 cass. 95.0
1500 cass. 82
55
45
Wax
85.0
4500 cass. 95.0
1500 cass. 100
100
77
Cleaning Waxsol
88.0
6 cycles* N/A
N/A
100
100
45
Cleaning ethanol
88.0
6 cycles* N/A
N/A
78
51
45