5 cleaning protocol, Cleaning protocol, Quick clean) – Leica Biosystems PELORIS_PELORIS II User Manual
Page 134

Reference
Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011
134
8.2.5 Cleaning Protocol
Quick Clean
Pre-defined protocol “Quick Clean“.
Bottle configuration: one bottle each of cleaning solvent and cleaning alcohol.
Never run a cleaning protocol while tissue is in the retort as the dry step will damage the
tissue.
Step Reagent group
Time (min) Temp (°C) P/V
Stirrer
Drip time
(s)
1
Cleaning solvents
12
75
Ambient
High
10
2
Cleaning alcohols
6
55
Ambient
High
10
3
Dry step
12
80
N/A
Off
N/A
Processing time
0:34:00
See also other documents in the category Leica Biosystems Equipment:
- SM2010 R (46 pages)
- CM305 S (66 pages)
- CE Knife holder (14 pages)
- CM1950 (78 pages)
- VT1200/VT1200 S (62 pages)
- VT1000 S (42 pages)
- HI1220 (18 pages)
- HI1210 (18 pages)
- CM1850 UV (56 pages)
- CM1850 UV (54 pages)
- CM1510-S (56 pages)
- CM3600 XP (114 pages)
- CM1100 (36 pages)
- CM1860/CM1860 UV (98 pages)
- Premium blade holder (14 pages)
- CM1520 (92 pages)
- VT1000 A (49 pages)
- VT1000 P (48 pages)
- SP2600 (51 pages)
- SM2500 (76 pages)
- SM2400 (8 pages)
- SM2000R (22 pages)
- EG1160 (32 pages)
- EG1150 C (20 pages)
- EG1150 H (42 pages)
- EG1130 (2 pages)
- EG1120 (2 pages)
- EG F (18 pages)
- ASP300 (70 pages)
- TP1020 (102 pages)
- ASP200 S (88 pages)
- ASP300 S (88 pages)
- ASP6025 (178 pages)
- LN22 (44 pages)
- RM CoolClamp (20 pages)
- RM2265 (88 pages)
- RM2255 (82 pages)
- RM2245 (78 pages)
- RM2235 (78 pages)
- RM2125 RTS (60 pages)
- SP9000 (30 pages)
- SP1600 (10 pages)
- ST4040 (50 pages)
- ST5020 (88 pages)