Leica Biosystems PELORIS_PELORIS II User Manual
Page 136

Reference
Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011
136
Xylene-free Configuration
This is the suggested station configuration when running xylene-free protocols (see 8.2.4 Xylene-
free Protocols). See note below for special conditions for initial bottle setup.
*All alcohol dehydrants after fixative should be labeled and named in the software “85% Ethanol”. However, if
setting up bottles on a new instrument or replacing all the ethanol bottles at once, the first two bottles should
be filled with 70% ethanol (but still be named “85% Ethanol”). As operation continues and you are prompted
to replace degraded bottles, replace with 85% ethanol. Other, initially higher-concentration bottles will have
degraded, so there will always be at least one bottle at a low concentration.
Station
Reagent type
Bottle 1
Formalin
Bottle 2
Formalin
Bottle 3
85% Ethanol*
Bottle 4
85% Ethanol*
Bottle 5
85% Ethanol
Bottle 6
80/20 Ethanol/IPA
Bottle 7
80/20 Ethanol/IPA
Bottle 8
80/20 Ethanol/IPA
Bottle 9
IPA
Bottle 10
IPA
Bottle 11
IPA
Bottle 12
IPA
Bottle 13
Waxsol™
Bottle 14
Waxsol™
Bottle 15
Cleaning ethanol
Bottle 16
Cleaning ethanol
Wax 1
Wax
Wax 2
Wax
Wax 3
Wax
Wax 4
Wax