Reagent Setup
Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011
82
Figure 45. Reagent types screen, temperature thresholds view, using recommended reagent types
for xylene-free processing
Supervisors can edit the default concentration, purity thresholds, and temperature thresholds for all
active reagent types.
You edit a type’s attributes by tapping the appropriate table cell then entering the required value
using the resultant keypad. The attributes update immediately and are applied to all reagent
stations and protocols that use the reagent type. The changes will not affect running protocols.
Lowering temperature thresholds may make protocol steps invalid. You must lower the step
temperature to comply with the new reagent threshold before you can load or run the protocol.
Leica Microsystems does not recommend you lower reagent concentration thresholds. Raising
thresholds can help to correct poor quality processing due to impure reagents.
If you no longer use a reagent in the active list you can remove it to the dormant list, to keep the
active list as small and easy to manage as possible. Select the reagent type by tapping in its name
cell, then tap Remove reagent.
List of active reagent
types
Purity thresholds
and Temperature
thresholds buttons to
switch views
Selected reagent type
Add reagent button:
bring dormant reagent
onto the active list
Remove reagent
button: remove
selected reagent from
the active to the
dormant list
Max. temperatures:
tap cells to change
Defaults/Concentration:
tap cells to change