4 poor processing – incorrect protocol, Poor processing – incorrect protocol – Leica Biosystems PELORIS_PELORIS II User Manual
Page 144

Troubleshooting
Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011
144
9.2.4 Poor Processing – Incorrect Protocol
Poor processing – incorrect
protocol
Was
the original
protocol too
long?
Tissue is over-
processed.
Reprocessing will not
help.
Was
the original
protocol too
short?
Tissue is under processed. Reprocess. NB: Do not use the Peloris clean
cycle to reprocess the tissue as the drying step will cause the tissue to dry
out.
Is it
large robust
tissue?
Is it fatty
tissue?
Try a longer protocol with
less heat and agitation.
Try a longer, protocol with
more heat and efficient
agitation.
Try a defatting step of 50/50
xylene and ethanol. Put step
between ethanols and xylene,
ensuring that there is a 100%
ethanol step after defat,
before xylene.
Is tissue
processed
adequately?
Yes
Yes
Yes
Yes
No
Was wax
infiltration
incomplete?
If tissue is dehydrated and cleared
there is no need for reprocessing.
Re-infiltrate tissue with wax only.
Yes
No
Adequately processed
tissue
Yes
Try:
·
soaking the block in a softening agent, such
as fabric softener, before cutting.
·
Cutting thinner sections.
·
Warming the block slightly before sectioning
No
Contact customer
support
Is it xylene
processing?
No
No
Is it
large delicate
tissue?
No
No
Yes
No