Language and Unit Settings
You can view the instrument’s language and temperature and pressure unit settings in the
Instrument settings screen. Contact your service representative to change either setting.
6.2.2 Device Settings
The Device settings screen (Control menu, Device settings) allows users to control display
settings and lock the screen for cleaning. Supervisors can configure instrument sounds and enable
and disable “power off” alarms for the external alarm connectors.
Sound Settings
Supervisors can select from a range of sounds to be heard for particular types of event, or have no
sound for these events. For example a sound can be set for key touches on the touch-screen, or to
accompany a prompt dialog box or warning. There is also volume control for all instrument sounds.
To change the sound associated with an event type, tap the cell beside the event type in the
Sound settings table. Select a sound file in the Select sound file dialog box then tap OK.
Select
To hear a sound in the Select sound file select it then tap Play.
To change sound volume tap the cell beside Master volume in the Sound settings table.
Set a value between 100 (loudest) and 0 (silent).
Touch-screen Settings
All users can change the brightness of the touch-screen on the Device settings screen. Tap the
the cell beside Brightness in the Touch screen settings table then enter the required value from
0 (dimmest) to 100 (brightest). The default value is 70%.
Alarm Power Off Settings
Use the alarm power off settings on the Device settings screen to control the behavior of the
external LOCAL ALARM and REMOTE ALARM (see 2.7 Alarm Connections) in the event of a power
failure. If the “power off” state is set to Enabled the alarm will activate (sound) if there is a power
failure.
To change the setting tap the cell beside Local alarm or Remote alarm in the Alarm power
off settings table then OK in the dialog box.
You require supervisor level access to alter these settings.