beautypg.com

4 recommended reagents, Recommended reagents – Leica Biosystems PELORIS_PELORIS II User Manual

Page 78

background image

Reagent Setup

Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011

78

5.1.4 Recommended Reagents

While each laboratory is responsible for its choice of reagents and wax, using reagents other than

those recommended for Peloris can result in poor processing or compromise instrument reliability.

Compliance with the following guidelines will ensure superior tissue processing.

The use of alternative reagents must be validated by the end user according to their local or

regional accreditation requirements.

Adequate fixation of tissue samples is required for optimum results and should be achieved prior to

placement on Peloris or incorporated into the fixative step of the protocol. The following reagents

have been validated for use on Peloris.

Fixatives

10% Neutral Buffered Formalin (NBF)

Alcohols

Histological grade ethanol

Reagent (grade) ethanol

Absolute ethanol

Isopropyl alcohol (dehydrant)

Denatured ethanol is acceptable if:

It is at least 99% ethanol, and

It is denatured with methanol and/or isopropyl alcohol (IPA)

Ethanol containing acetone must not be used.

Clearing Agents

Xylene is the recommended clearing agent.

Xylene-free processing requires the use of isopropyl alcohol as the clearing agent. It has been

fully tested and validated.

Paraffin

Use histological wax for histology

Parablocks™

Paraplast®

Cleaning Agents

Use ethanol for the cleaning alcohol solution.

In xylene processing mode, we recommend xylene as the cleaning solvent.