Xylene 2 hour, Xylene 4 hour – Leica Biosystems PELORIS_PELORIS II User Manual
Page 128

Reference
Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011
128
Xylene 2 Hour
Pre-defined protocol “Factory 2hr Xylene Standard“
Carryover setting in pre-defined protocol: 75
Xylene 4 Hour
Pre-defined protocol “Factory 4hr Xylene Standard“
Carryover setting in pre-defined protocol: 50
Step Reagent type
Reagent
group
Time
(min)
Temp
(°C)
P/V
Stirrer
Drip time
(s)
1
Formalin
Fixatives
1
Ambient
Ambient
Medium
10
2
Ethanol
Dehydrants
1
Ambient
Ambient
Medium
10
3
Ethanol
Dehydrants
1
Ambient
Ambient
Medium
10
4
Ethanol
Dehydrants
1
Ambient
Ambient
Medium
10
5
Ethanol
Dehydrants
1
Ambient
Ambient
Medium
10
6
Ethanol
Dehydrants
11
45
Ambient
Medium
10
7
Ethanol
Dehydrants
30
45
Ambient
Medium
10
8
Xylene
Clearers
1
Ambient
Ambient
Medium
10
9
Xylene
Clearers
1
Ambient
Ambient
Medium
10
10
Xylene
Clearers
28
45
Ambient
Medium
10
11
Paraffin wax
Wax
5
65
Vacuum
Medium
10
12
Paraffin wax
Wax
5
65
Vacuum
Medium
10
13
Paraffin wax
Wax
20
65
Vacuum
Medium
10
Processing time
2:14:00
Step Reagent type
Reagent
group
Time
(min)
Temp
(°C)
P/V
Stirrer
Drip time
(s)
1
Formalin
Fixatives
10
45
Ambient
Medium
10
2
Ethanol
Dehydrants
1
Ambient
Ambient
Medium
10
3
Ethanol
Dehydrants
1
Ambient
Ambient
Medium
10
4
Ethanol
Dehydrants
1
Ambient
Ambient
Medium
10
5
Ethanol
Dehydrants
20
45
Ambient
Medium
10
6
Ethanol
Dehydrants
20
45
Ambient
Medium
10
7
Ethanol
Dehydrants
45
45
Ambient
Medium
10
8
Xylene
Clearers
1
Ambient
Ambient
Medium
10
9
Xylene
Clearers
10
45
Ambient
Medium
10
10
Xylene
Clearers
45
45
Ambient
Medium
10
11
Paraffin wax
Wax
10
65
Vacuum
Medium
10
12
Paraffin wax
Wax
10
65
Vacuum
Medium
10
13
Paraffin wax
Wax
40
65
Vacuum
Medium
10
Processing time
4:02:00