3 xylene protocols, Xylene protocols – Leica Biosystems PELORIS_PELORIS II User Manual
Page 127

Reference
Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011
127
The pre-defined protocols are:
Factory 1hr Xylene Standard (see Xylene 1 Hour)
Factory 2hr Xylene Standard (see Xylene 2 Hour)
Factory 4hr Xylene Standard (see Xylene 4 Hour)
Factory 8hr Xylene Standard (see Xylene 8 Hour)
Factory 12hr Xylene Standard (see Xylene 12 Hour)
Factory 1 hr Xylene Free (see Xylene-free 1 Hour)
Factory 2 hr Xylene Free (see Xylene-free 2 Hour)
Factory 4 hr Xylene Free (see Xylene-free 4 Hour)
Factory 8 hr Xylene Free (see Xylene-free 8 Hour)
Factory 12 hr Xylene Free (see Xylene-free 12 Hour)
Quick Clean (see Quick Clean)
8.2.3 Xylene Protocols
All the xylene protocols below are designed for use with the xylene bottle configuration in 8.3
Xylene 1 Hour
Pre-defined protocol “Factory 1hr Xylene Standard“
Carryover setting in pre-defined protocol: 100
Step Reagent type
Reagent
group
Time
(min)
Temp
(°C)
P/V
Stirrer
Drip time
(s)
1
Formalin
Fixatives
1
Ambient
Ambient
Medium
10
2
Ethanol
Dehydrants
1
Ambient
Ambient
Medium
10
3
Ethanol
Dehydrants
1
Ambient
Ambient
Medium
10
4
Ethanol
Dehydrants
1
Ambient
Ambient
Medium
10
5
Ethanol
Dehydrants
1
Ambient
Ambient
Medium
10
6
Ethanol
Dehydrants
1
Ambient
Ambient. Medium
10
7
Ethanol
Dehydrants
18
45
Ambient
Medium
10
8
Xylene
Clearers
1
Ambient
Ambient
Medium
10
9
Xylene
Clearers
1
Ambient
Ambient
Medium
10
10
Xylene
Clearers
14
45
Ambient
Medium
10
11
Paraffin wax
Wax
2
65
Vacuum
Medium
10
12
Paraffin wax
Wax
1
65
Vacuum
Medium
10
13
Paraffin wax
Wax
14
65
Vacuum
Medium
10
Processing time