How to schedule a protocol run – Leica Biosystems PELORIS_PELORIS II User Manual
Page 51

Running Protocols
Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011
51
How To Schedule a Protocol Run
When the Scheduling dialog box opens the retort’s default end time is shown.
If the default end time is set to ASAP the protocol is scheduled to end at the earliest possible
time.
If the default setting is a specific time of day, the dialog box defaults to that time on the
following day.
You have four options:
1. Accept the default values as they are.
Edit the protocol for the run if required (see 3.4.2 Editing the Protocol for a Single Run) then
tap Start to begin the run.
2. Keep the default time but change the day the protocol will run.
Tap the + day or – day button to keep the default end time but change the day.
If the default setting is a specific time of day the dialog box opens with the run set to the
following day. To run the protocol on the current day tap the – day button once. Depending on
the time you do this the new end time may not be possible. If so, the Predicted end time is
highlighted red.
3. Start the run as soon as possible.
Tap the ASAP button.
The protocol will normally start immediately. However, if there is a reagent clash caused by a
protocol running in the other retort the actual start of the protocol may be delayed (see 3.6.1
Delayed End Times and Initial Fills).
4. Set a new end time.
Tap the Edit end time button to open the Run protocol – edit end time dialog box (see
the next section for instructions).