JULABO SC 10000w Professional Series Recirculating Coolers User Manual
Page 70

Remote control
70
in commands:
Asking for parameters or temperature values to be displayed.
Command
Parameter
Response of recirculating cooler
version
none
Number of software version (V X.xx)
status
none
Status message, error message (see page 72)
in_pv_00
none
Actual bath temperature.
in_pv_01
none
Heating power being used (%).
in_pv_02
none
Temperature value registered by the external Pt100 sensor.
in_pv_03
none
Temperature value registered by the safety sensor.
in_pv_04
none
Setpoint temperature of the excess temperature protection
in_sp_00
none
Working temperature „SETP 1“
in_sp_01
none
Working temperature „SETP 2“
in_sp_02
none
Working temperature „SETP 3“
in_sp_03
none
High temperature warning limit
„OVERTEMP“
in_sp_04
none
Low temperature warning limit
„SUBTEMP“
in_sp_05
none
Setpoint temperature of the external programmer
(REG+E-PROG) .
in_sp_07
none
Adjusted pump stage
in_sp_08
none
Value of a flowrate sensor connected to the E-Prog input
in_par_00
none
Temperature difference between working sensor and safety
sensor
in_par_01
none
Te - Time constant of the external bath.
in_par_02
none
Si - Internal slope
in_par_03
none
Ti - Time constant of the internal bath.
in_par_04
none
CoSpeed - Band limit (max. difference between the
temperatures in the internal bath and external system).
in_par_05
none
Factor pk/ph0:
Ratio of max. cooling capacity versus max. heating capacity
in_par_06
none
Xp control parameter of the internal controller.
in_par_07
none
Tn control parameter of the internal controller.
in_par_08
none
Tv control parameter of the internal controller.
in_par_09
none
Xp control parameter of the cascade controller.
in_par_10
none
Proportional portion of the cascade controller.