JULABO SC 10000w Professional Series Recirculating Coolers User Manual
Page 68

Remote control
68
To ensure a safe data transfer, the time gap between two commands should
be at least 250 ms.
The circulator automatically responds to an in command with a data string
followed by a LF (Line Feed). The next command should only be sent after
10 ms.
The out commands are valid only in remote control mode.
Examples:
Command to set the working temperature >Setpoint1< to 55.5 °C
out_sp_00
55.5
Command to ask for the working temperature >Setpoint1<
in_sp_00
Response from the circulator:
55.5
LF
12.3. List of commands
out commands:
Setting temperature values or parameters.
Command
Parameter
Response of recirculating cooler
out_mode_01
0
Use working temperature > SETP 1<
out_mode_01
1
Use working temperature > SETP 2<
out_mode_01
2
Use working temperature > SETP 3<
out_mode_02
0
Selftuning „OFF“.
Temperature control by using the stored parameters.
out_mode_02 1
Selftuning
„ONCE“
Single selftuning of controlled system after the next start.
out_mode_02 2
Selftuning
„ALWAYS“
Continual selftuning of controlled system whenever a new
setpoint is to be reached.
out_mode_03
0
Set external programmer input to voltage.
Voltage
0 V ... 10 V
out_mode_03
1
Set external programmer input to current.
Current
0 mA ... 20 mA
out_mode_04
0
Temperature control of internal bath.
out_mode_04
1
External control with Pt100 sensor.
out_mode_05
0
Stop the unit = R –OFF-.
out_mode_05
1
Start the unit.
out_mode_08
0
Set the control dynamics - aperiodic