JULABO SC 10000w Professional Series Recirculating Coolers User Manual
Page 35
Recirculating Cooler
35
1.
Press the
button until the submenu >CONTROL< is displayed.
2.
Select the parameter with the keys
(INT / EXT).
3.
Press enter
to store the selected parameter.
Continue: Press
or
Notice:
Place the external sensor into the bath medium and securely fix the sensor.
Pt100
IMPORTANT:
Additional measures for external temperature control.
For external control and temperature measurement an external
Pt100 sensor must be connected to the socket (17) on the rear of the
recirculating cooler.
Suggested adjustments for external temperature control:
BAND HIGH / LOW and INTERN MAX / MIN
see chapter > LIMITS < on page 47.
Sensor calibration of the external Pt100 sensor is carried out in the
>MENU / ADJUST<, in the submenu >ATC SEN / EXT<
(see page 53).
Pt100
M+R
Accessory: Pt100 external sensor
Order No.
Description
Material
Cable
8981003
200x6 mm Ø,
stainless steel
1.5 m
8981005
200x6 mm Ø,
glass
1.5 m
8981006
20x2 mm Ø,
stainless steel
1.5 m
8981010
300x6 mm Ø,
stainless steel
1.5 m
8981015
300x6 mm Ø,
stainless steel / PTFE coated
3 m
8981013
600x6 mm Ø,
stainless steel / PTFE coated
3 m
8981016
900x6 mm Ø,
stainless steel / PTFE coated
3 m
8981014
1200x6 mm Ø,
stainless steel / PTFE coated
3 m
8981103 Extension cable for Pt100 sensor
3.5 m
8981020 M+R in-line Pt100 sensor
The M+R in-line Pt100 sensor is a flow sensor and can be installed loop
circuit