beautypg.com

JULABO SC 10000w Professional Series Recirculating Coolers User Manual

Page 35

background image

Recirculating Cooler

35

1.

Press the

button until the submenu >CONTROL< is displayed.

2.

Select the parameter with the keys

(INT / EXT).

3.

Press enter

to store the selected parameter.

Continue: Press

or

Notice:

Place the external sensor into the bath medium and securely fix the sensor.

Pt100

IMPORTANT:

Additional measures for external temperature control.

For external control and temperature measurement an external
Pt100 sensor must be connected to the socket (17) on the rear of the
recirculating cooler.

Suggested adjustments for external temperature control:
BAND HIGH / LOW and INTERN MAX / MIN
see chapter > LIMITS < on page 47.

Sensor calibration of the external Pt100 sensor is carried out in the
>MENU / ADJUST<, in the submenu >ATC SEN / EXT<
(see page 53).

Pt100

M+R

Accessory: Pt100 external sensor

Order No.

Description

Material

Cable

8981003

200x6 mm Ø,

stainless steel

1.5 m

8981005

200x6 mm Ø,

glass

1.5 m

8981006

20x2 mm Ø,

stainless steel

1.5 m

8981010

300x6 mm Ø,

stainless steel

1.5 m

8981015

300x6 mm Ø,

stainless steel / PTFE coated

3 m

8981013

600x6 mm Ø,

stainless steel / PTFE coated

3 m

8981016

900x6 mm Ø,

stainless steel / PTFE coated

3 m

8981014

1200x6 mm Ø,

stainless steel / PTFE coated

3 m

8981103 Extension cable for Pt100 sensor

3.5 m

8981020 M+R in-line Pt100 sensor

The M+R in-line Pt100 sensor is a flow sensor and can be installed loop
circuit