Troubleshooting guide / error messages – JULABO SC 10000w Professional Series Recirculating Coolers User Manual
Page 63
Recirculating Cooler
63
10. Troubleshooting guide / Error messages
XXXXX
ALARM
CODE 01
+
Alarm
with a complete shutdown of the unit
Whenever the microprocessor electronics registers a failure, a complete
shutdown of the heater and circulating pump is performed. The alarm light
„
“ illuminates and a continuous signal tone sounds.
The VFD COMFORT-DISPLAY indicates the cause for the alarm in form of a
code.
XXXXX
WARNING
CODE 40
Warning
without a complete shutdown of the unit
The VFD COMFORT-DISPLAY indicates the cause for the warning in form of
a code and an acoustic signal sounds in regular intervals.
These messages appear every 10 seconds.
Press enter
to quit the audible signal.
ALARM
CODE 01
The recirculating cooler is operated without bath fluid, or the liquid level is
insufficient. Replenish the bath tank with the bath fluid.
Tube breakage has occured (insufficient filling level due to excessive bath
fluid pumped out). Replace the tubing and replenish the bath tank with the
bath fluid.
The float is defect (e. g., because damaged in transit). Repair by
authorized JULABO service personnel.
Excess temperature sensors defect.
The excess temperature value lies below the working temperature
setpoint. Set the excess temperature to a higher value.
WARNING
CODE 03
ALARM
CODE 03
Excess temperature warning
or
Excess temperature alarm
Warning type:
Set to>Warning< or >Alarm< (see page 31)
WARNING
CODE 04
ALARM
CODE 04
Low temperature warning
or
Low temperature alarm.
Warning type:
Set to>Warning< or >Alarm< (see page 31)
ALARM
CODE 05
Cable of the working temperature sensor interrupted or short-circuited.
ALARM
CODE 06
Defect of the working or excess temperature sensor.
Working temperature and excess temperature sensors report a
temperature difference of more than 35 K.
ALARM
CODE 07
Other errors (I
2
C-BUS errors)