beautypg.com

Setting the temperatures – JULABO SC 10000w Professional Series Recirculating Coolers User Manual

Page 27

background image

Recirculating Cooler

27

7.

Setting the temperatures

Factory settings:

SETP 1 25 °C

SETP 2 37 °C

SETP 3 70 °C

Press the

key to call up the menu for temperature selection.

3 different working temperatures are adjustable. Their values are freely
selectable within the operating temperature range.

This setting may be carried out with the recirculating cooler being in the
Start or Stop condition!

Example: Setting working temperature "SETP

oint 3"

XXXXX

SETP 3
70.0


XXXXX

SETP 3
85.0



1.

Press the

key until the desired menu window is indicated on the

VFD COMFORT-DISPLAY
Example: SETP 3 / 70.0 °C (last digit blinks)

2.

Change the value to 85 °C.

Use the cursor keys to move left or right on the display until the
numeral you wish to change blinks.

Use the increase/decrease arrows

to change the selected

numeral (-, 0, 1, 2, 3, ... 9).
see example left: SETP 3 / 85.0.

3.

Press enter

to store the value.

In the >Start< condition this value is immediately used for controlling the
working temperature.
The indication on the VFD COMFORT-DISPLAY is updated.
The heater control indicator blinks.

Notice:

See SETMAX and SETMIN in chapter 9.5. MENU LIMITS

Example: Selecting the working temperature

SETP 1
SETP 2
SETP 3

1.

Press the

key until the desired menu item is indicated on the VFD

COMFORT-DISPLAY.

2.

Press enter

.

The recirculating cooler uses the new working temperature value for
temperature control.