Setting the temperatures – JULABO SC 10000w Professional Series Recirculating Coolers User Manual
Page 27

Recirculating Cooler
27
7.
Setting the temperatures
Factory settings:
SETP 1 25 °C
SETP 2 37 °C
SETP 3 70 °C
Press the
key to call up the menu for temperature selection.
3 different working temperatures are adjustable. Their values are freely
selectable within the operating temperature range.
This setting may be carried out with the recirculating cooler being in the
Start or Stop condition!
Example: Setting working temperature "SETP
oint 3"
XXXXX
SETP 3
70.0
XXXXX
SETP 3
85.0
1.
Press the
key until the desired menu window is indicated on the
VFD COMFORT-DISPLAY
Example: SETP 3 / 70.0 °C (last digit blinks)
2.
Change the value to 85 °C.
Use the cursor keys to move left or right on the display until the
numeral you wish to change blinks.
Use the increase/decrease arrows
to change the selected
numeral (-, 0, 1, 2, 3, ... 9).
see example left: SETP 3 / 85.0.
3.
Press enter
to store the value.
In the >Start< condition this value is immediately used for controlling the
working temperature.
The indication on the VFD COMFORT-DISPLAY is updated.
The heater control indicator blinks.
Notice:
See SETMAX and SETMIN in chapter 9.5. MENU LIMITS
Example: Selecting the working temperature
SETP 1
SETP 2
SETP 3
1.
Press the
key until the desired menu item is indicated on the VFD
COMFORT-DISPLAY.
2.
Press enter
.
The recirculating cooler uses the new working temperature value for
temperature control.