Menu functions

34

9.2. MENU CONTROL – Control parameters

CONTROL: Press enter

to switch to menu level 2

>CONTROL<

(INT /

EXT

)

>SELFTUN<

(ALWAYS / OFF / ONCE)

>DYNINT<

(APER / NORM)

>XP INT<

(0.1 ... 99.9)

>TN INT<

(3 ... 9999)

>TV INT<

(0 ... 999)

or

>CONTROL<

(

INT

/ EXT)

>SELFTUN<

(ALWAYS / OFF / ONCE)

>COSPEED<

(0 ... 5)

>XP EXT<

(0.1 ... 99.9)

>TN EXT<

(3 ... 9999)

> TV EXT

(0 ... 999)

>XPU EXT<

(0.1 ... 99.9)





9.2.1.

CONTROL – internal / external control

Factory setting:
INT

The recirculating cooler is conceived for internal and external
temperature control. Switching is carried out in this submenu.
Depending on what is set, only the respective set of parameters is
indicated.

Possible parameters:

INT

internal temperature control

EXT

external temperature control with external Pt100 sensor

The control type can only be adjusted in the -OFF- condition