9.2.2. SELFTUNING

XXXXX

SELFTUN
ALWAYS

Factory setting:
ONCE

Selftuning:

When performing a selftuning for the controlled system (temperature
application system), the control parameters Xp, Tn and Tv are
automatically determined and stored.

Possible parameters:

OFF

- no selftuning

The control parameters ascertained during the last identification are
used for control purposes.

ONCE

- single selftuning (factory setting)

The instrument performs a single selftuning of the controlled system
after each start with the start/stop key

or after receiving a start

command via the interface.

ALWAYS

- continual selftuning

The instrument performs a selftuning of the controlled system
whenever a new setpoint is to be reached.
Use this setting only when the temperature application system
changes permanently.

1.

Press the

button until the submenu > SELFTUN < is displayed.

2.

Select the parameter with the keys

.

(ALWAYS / OFF / ONCE)

3.

Press enter

to store the selected parameter.

Continue: Press

or